86
Chapter 5: Troubleshooting
Troubleshooting the Agilent 1680A,AD-series
Possible problem with
logic analyzer cables. Do
the cable test in Chapter
5 on suspect pod.
Does cable
test pass?
Swap suspect probe
tip assembly with
known good one.
Does cable
test pass?
Swap suspect cable
assembly with known
good one.
Does cable
test pass?
Replace logic analyzer
acquisition board.
Replace
defective cable.
Replace defective
probe tip assembly.
The logic analyzer
board if functioning
properly.
No
Yes
No
Yes
No
Yes
7
1
Summary of Contents for 1680 series
Page 13: ...13 Chapter 1 General Information Dimensions 1680A AD Series 1690A AD Series...
Page 74: ...74 Chapter 3 Testing Performance Performance Test Record...
Page 142: ...142 Chapter 6 Replacing Assemblies 1690A AD series disassembly assembly...
Page 172: ...172 Chapter 8 Theory of Operation Self Tests Descriptions...
Page 174: ......