96
Chapter 5: Troubleshooting
Troubleshooting the Agilent 1690A,AD-series
Reconnect the
power cord.
Are the instrument
fans running?
Suspect the power supply.
Perform "To test the
power supply voltages"
in Chapter 5.
Do all test
points pass?
Ensure all instrument
cables are connected
and properly seated.
Replace the
power supply.
Yes
No
Yes
No
Are the power
supply fans
running?
Yes
No
1
2
Ensure the power cord
is properly connected.
Is the power cord
connected?
Yes
No