33
Chapter 3: Testing Logic Analyzer Performance
Connect the Test Equipment
Connect the Test Equipment
Connect the 16910/11A Logic Analyzer Pod to the 8133A Pulse
Generator
1
Connect a Transition Time Converter (if required—see page 28) to each of
the four outputs of the 8133A pulse generator except Channel 1 OUTPUT.
2
Connect the SMA/Flying Lead test connector (see “To Assemble the SMA/
Flying Lead Test Connectors” on page 22)
with
the 50 ohm terminator to
the Transition Time Converter at the 8133A pulse generator Channel 1
OUTPUT. (If Transition Time Converters are not required, connect the
SMA/Flying Lead test connector directly to the pulse generator output.)
3
Connect the two SMA/Flying Lead test connectors
without
50 ohm
terminators to the Transition Time Converters at the 8133A pulse
generator Channel 2 OUTPUT and Channel 2 OUTPUT. (If Transition Time
Converters are not required, connect the SMA/Flying Lead test connectors
directly to the pulse generator outputs.)
4
Connect an E5383A Flying Lead Probe Set to Pod 1 of the 16910/11A logic
analyzer.
Summary of Contents for 16900 Series
Page 3: ...3 Chapter The 16910A Logic Analyzer The 16911A Logic Analyzer...
Page 8: ...8 Contents...
Page 14: ...14 Chapter 1 General Information...
Page 18: ...18 Chapter 2 Preparing for Use...
Page 61: ...61 4 Calibrating This chapter gives you instructions for calibrating the logic analyzer...
Page 65: ...65 Chapter 5 Troubleshooting Troubleshooting Flowchart 1...
Page 66: ...66 Chapter 5 Troubleshooting Troubleshooting Flowchart 2...
Page 82: ...82 Chapter 5 Troubleshooting To test the cables 18 Return to the troubleshooting flow chart...
Page 94: ...94 Chapter 7 Replaceable Parts 16910A Exploded View Exploded view of the 16910A logic analyzer...
Page 95: ...95 Chapter 7 Replaceable Parts 16911A Exploded View Exploded view of the 16911A logic analyzer...
Page 96: ...96 Chapter 7 Replaceable Parts...
Page 102: ...102 Index...