71
Chapter 5: Troubleshooting
To Assemble the 2 x 9 Test Connectors
To Assemble the 2 x 9 Test Connectors
The 2 x 9 test connectors are used to connect all 16 channels and the clock of the
logic analyzer to the pulse generator so you can test the flying lead probe and
cables. (See “To test the cables” on page 73.)
Materials Required
Build two test connectors using SMA connectors and 2-by-9 sections of pin strip.
1
Prepare the pin strip header:
a
Cut or cleanly break two 2 x 9 sections from the pin strip.
b
Solder a jumper wire to all nine pins on one side of the pin strip.
c
Solder a jumper wire to all nine pins on the other side of the pin strip.
d
Solder two resistors to the pin strip, one at each end between the end
Material
Critical Specification
Recommended Model/Part
Pin Strip Header
(Qty 1, which will be separated)
.100" x .100" Pin Strip Header,
straight, pin length .230", two rows,
.120" solder tails, 2 x 40 contacts
3M 2380-6121TN or similar 2- row
with 0.1” pin spacing
Jumper wire
<6 inches, approximately 22 gauge
Resistor, 100 ohm 1%
(Qty 4)
SMA Board Mount Connector
(Qty 2)
Johnson 142-0701-801
(see www.johnsoncomponents.com)
Summary of Contents for 16900 Series
Page 3: ...3 Chapter The 16910A Logic Analyzer The 16911A Logic Analyzer...
Page 8: ...8 Contents...
Page 14: ...14 Chapter 1 General Information...
Page 18: ...18 Chapter 2 Preparing for Use...
Page 61: ...61 4 Calibrating This chapter gives you instructions for calibrating the logic analyzer...
Page 65: ...65 Chapter 5 Troubleshooting Troubleshooting Flowchart 1...
Page 66: ...66 Chapter 5 Troubleshooting Troubleshooting Flowchart 2...
Page 82: ...82 Chapter 5 Troubleshooting To test the cables 18 Return to the troubleshooting flow chart...
Page 94: ...94 Chapter 7 Replaceable Parts 16910A Exploded View Exploded view of the 16910A logic analyzer...
Page 95: ...95 Chapter 7 Replaceable Parts 16911A Exploded View Exploded view of the 16911A logic analyzer...
Page 96: ...96 Chapter 7 Replaceable Parts...
Page 102: ...102 Index...