Chapter 3: Testing Logic Analyzer Performance
To ensure the 16910A or 16911A logic analyzer (also referred to as the module or
the card) is operating as specified, software tests (self-tests) and a manual
performance test are done. The logic analyzer is considered performance-verified
if all of the software tests and the manual performance test have passed. The
procedures in this chapter indicate what constitutes a “Pass” status for each of
the tests.
Test Strategy
This chapter describes the module being tested in an Agilent Technologies
16900A-series mainframe.
Only specified parameters are tested. Specifications are listed on page 11. The
test conditions defined in this procedure ensure that the specified parameter is
as good as or better than specification. No attempt is made to determine
performance which is better than specification. Not all channels of the logic
analyzer will be tested; rather a sample of channels is tested. The calibration
laboratory may choose to elaborate on these tests and test all channels at their
discretion.
One-card Module.
To perform a complete test on a one-card module, start at
the beginning of the chapter and follow each procedure.
Multi-card Module.
To perform a complete test on a multi-card module,
perform the self-tests (page 30) with the cards connected. Then, remove the
multi-card module from the mainframe and configure each card as a one-card
module. Install the one-card modules into the mainframe and perform the
performance verification tests on each card. When the tests are complete,
remove the one-card modules, reconfigure them into their original multi-card
module configuration, reinstall it into the mainframe and perform the self-tests
again. These steps are necessary to ensure that the clocks are tested on each
module.
For removal instructions see Chapter 6, “Replacing Assemblies,” beginning on
page 83. For installation and configuration instructions see Chapter 2, “Preparing
for Use,” beginning on page 15.
Test Interval
Test the performance of the module against specifications at two-year intervals.
Test Record Description
A Performance Test Record for recording the results of each procedure is located
at the end of this chapter.
Summary of Contents for 16900 Series
Page 3: ...3 Chapter The 16910A Logic Analyzer The 16911A Logic Analyzer...
Page 8: ...8 Contents...
Page 14: ...14 Chapter 1 General Information...
Page 18: ...18 Chapter 2 Preparing for Use...
Page 61: ...61 4 Calibrating This chapter gives you instructions for calibrating the logic analyzer...
Page 65: ...65 Chapter 5 Troubleshooting Troubleshooting Flowchart 1...
Page 66: ...66 Chapter 5 Troubleshooting Troubleshooting Flowchart 2...
Page 82: ...82 Chapter 5 Troubleshooting To test the cables 18 Return to the troubleshooting flow chart...
Page 94: ...94 Chapter 7 Replaceable Parts 16910A Exploded View Exploded view of the 16910A logic analyzer...
Page 95: ...95 Chapter 7 Replaceable Parts 16911A Exploded View Exploded view of the 16911A logic analyzer...
Page 96: ...96 Chapter 7 Replaceable Parts...
Page 102: ...102 Index...