28
Chapter 3: Testing Logic Analyzer Performance
Equipment Required
Equipment Required
The following equipment is required for the performance test procedure.
Equipment Required
Equipment
Critical Specification
Recommended Model/Part
Pulse Generator
≥
260 MHz, two channels, differential
outputs, 150-180 ps rise/fall time (if faster,
use transition time converters)
Agilent or HP 8133A option 003
150 ps Transition Time Converter
(Qty 3)
Required if pulse generator’s rise time is
less than 150 ps. (Pulse generator
conditions: Voffset=1V,
∆
V=500 mV.)
Required for 8133A opt. 003.
Agilent or HP 15435A
Oscilloscope
bandwidth
≥
1.5 GHz,
sampling rate
≥
8 GSa/s
Agilent or HP 54845A/B or similar
SMA Coax Cable
(Qty 2)
>18 GHz bandwidth
Agilent or HP 8120-4948
Flying Lead Probe Set with 5
ground leads
(Qty 2)
no substitute
Agilent or HP E5383A
Male BNC to Female SMA
adapters
(Qty 2)
Cambridge Products CP-AD507
(see www.cambridgeproducts.com)
SMA/Flying Lead test connectors,
(f) SMA to (f) SMA to Flying Lead
Probe
(Qty 3)
no substitute
See “To Assemble the SMA/Flying Lead
Test Connectors” on page 22
Summary of Contents for 16900 Series
Page 3: ...3 Chapter The 16910A Logic Analyzer The 16911A Logic Analyzer...
Page 8: ...8 Contents...
Page 14: ...14 Chapter 1 General Information...
Page 18: ...18 Chapter 2 Preparing for Use...
Page 61: ...61 4 Calibrating This chapter gives you instructions for calibrating the logic analyzer...
Page 65: ...65 Chapter 5 Troubleshooting Troubleshooting Flowchart 1...
Page 66: ...66 Chapter 5 Troubleshooting Troubleshooting Flowchart 2...
Page 82: ...82 Chapter 5 Troubleshooting To test the cables 18 Return to the troubleshooting flow chart...
Page 94: ...94 Chapter 7 Replaceable Parts 16910A Exploded View Exploded view of the 16910A logic analyzer...
Page 95: ...95 Chapter 7 Replaceable Parts 16911A Exploded View Exploded view of the 16911A logic analyzer...
Page 96: ...96 Chapter 7 Replaceable Parts...
Page 102: ...102 Index...