32
Chapter 3: Testing Logic Analyzer Performance
Set Up the Test Equipment
3
Set up the oscilloscope.
a
Set up the oscilloscope according to the following tables.
Oscilloscope Setup
Setup: Channel 1
Setup: Ch. 1 Probe
Setup: Channel 2
Setup: Ch. 2 Probe
On
Attenuation: 1.00:1
On
Attenuation: 1.00:1
Scale: 100 mV/div
Units: Volts
Scale: 100 mV/div
Units: Volts
Offset: 1 V
Attenuation Units: Ratio
Offset: 1 V
Attenuation Units: Ratio
Coupling: DC
External Gain: (n/a)
Coupling: DC
External Gain: (n/a)
Input: 50 ohm
Skew: (Set later. See page
37)
Input: 50 ohm
Skew: 0.0 seconds
External Offset: (n/a)
External Offset: (n/a)
Setup: Channel 3
Setup: Channel 4
Off
Off
Setup: Horizontal
Setup: Trigger
Setup: Acquisition
Setup: Display
Scale: 2 ns
Mode: Edge
Sampling Mode: Equiv. Time Waveforms: Connect dots
Position: 725 ps
Source: Channel 1
Memory Depth: Automatic
Persistence: Minimum
Reference: Center
Level: 1.00 V
Averaging: Enabled
Grid: On (and set intensity)
Delayed: not selected
Edge: Rising Edge
# of Averages: 4
Backlight Saver: as
preferred
Sweep: Auto
Measure: Markers
Mode: Manual placement
All else: (n/a)
Summary of Contents for 16900 Series
Page 3: ...3 Chapter The 16910A Logic Analyzer The 16911A Logic Analyzer...
Page 8: ...8 Contents...
Page 14: ...14 Chapter 1 General Information...
Page 18: ...18 Chapter 2 Preparing for Use...
Page 61: ...61 4 Calibrating This chapter gives you instructions for calibrating the logic analyzer...
Page 65: ...65 Chapter 5 Troubleshooting Troubleshooting Flowchart 1...
Page 66: ...66 Chapter 5 Troubleshooting Troubleshooting Flowchart 2...
Page 82: ...82 Chapter 5 Troubleshooting To test the cables 18 Return to the troubleshooting flow chart...
Page 94: ...94 Chapter 7 Replaceable Parts 16910A Exploded View Exploded view of the 16910A logic analyzer...
Page 95: ...95 Chapter 7 Replaceable Parts 16911A Exploded View Exploded view of the 16911A logic analyzer...
Page 96: ...96 Chapter 7 Replaceable Parts...
Page 102: ...102 Index...