45
Chapter 3: Testing Logic Analyzer Performance
Configure the Logic Analysis System
Run Eye Finder
6
Select the
Run
button in the Eye Finder window.
7
Ensure that an eye appears for each bit near the recommended starting
position. Depending on your test setup, the eye position may vary. Any
skew between channel 1 and channel 2 of your pulse generator will cause
the eye position to shift to the left or right in the Eye Finder display. A
shift of up to 0.5 ns should be considered normal.
The important point is that your Eye Finder display should look similar to
the picture below (although it may be shifted left or right), and Eye Finder
must be able to place the blue bars in the narrow eye. (The example below
shows Eye Finder in the 250 Mb/s mode.)
Summary of Contents for 16900 Series
Page 3: ...3 Chapter The 16910A Logic Analyzer The 16911A Logic Analyzer...
Page 8: ...8 Contents...
Page 14: ...14 Chapter 1 General Information...
Page 18: ...18 Chapter 2 Preparing for Use...
Page 61: ...61 4 Calibrating This chapter gives you instructions for calibrating the logic analyzer...
Page 65: ...65 Chapter 5 Troubleshooting Troubleshooting Flowchart 1...
Page 66: ...66 Chapter 5 Troubleshooting Troubleshooting Flowchart 2...
Page 82: ...82 Chapter 5 Troubleshooting To test the cables 18 Return to the troubleshooting flow chart...
Page 94: ...94 Chapter 7 Replaceable Parts 16910A Exploded View Exploded view of the 16910A logic analyzer...
Page 95: ...95 Chapter 7 Replaceable Parts 16911A Exploded View Exploded view of the 16911A logic analyzer...
Page 96: ...96 Chapter 7 Replaceable Parts...
Page 102: ...102 Index...