To remove and replace the front panel assembly
1
Perform previous procedures:
•
“To prepare the instrument for disassembly”
•
“To remove and replace the cover”
2
Disconnect the following cables:
•
USB cables from the motherboard.
•
Display cable from the display board (use care when
disconnecting to avoid damage).
•
Touch screen cable from the touch screen controller
board.
•
Keyboard cable from the keypad board.
•
Front panel cable from the backlight inverter board.
3
Using a Torx T10 screwdriver, remove 4 screws that
secure the shroud to the chassis.
4
Remove the shroud.
T-10
screws (4)
Shroud