Replacing Assemblies
6
16901A Logic Analysis System Service Guide
69
To remove and replace the tray assembly
1
Perform previous procedures:
•
“To prepare the instrument for disassembly”
•
“To remove and replace the cover”
•
“To remove and replace the front panel assembly”
2
Using a Torx T10 screwdriver, remove the 2 side screws
securing the tray assembly to the chassis.
3
Slide the tray assembly out far enough to disconnect the
hard drive and power supply cables. Use the release
tab(s) to help disconnect the power supply cables.
4
Slide the tray out of the chassis.
5
Reverse this procedure to replace the tray assembly.
T-10
screws (2)
Hard
drive
cables
Power supply
cables
Chassis