Replacing Assemblies
6
16901A Logic Analysis System Service Guide
61
To remove and replace the touch screen controller board
1
Perform previous procedures:
•
“To prepare the instrument for disassembly”
•
“To remove and replace the cover”
•
“To remove and replace the front panel assembly”
2
Disconnect the touch screen cable from the touch screen
controller board.
3
Using a Torx T10 screwdriver, remove the 2 screws
securing the board to the front panel board.
4
Reverse this procedure to install the touch screen
controller board.
Touch
screen
controller
board
T-10
screws
(2)
Touch screen
cable