6
Using E4410 E-Series Power Sensors
172
N1911A/1912A P-Series Power Meters User’s Guide
Now set the frequency of the signal you want to measure. The power
meter automatically selects the appropriate calibration factor.
5
Press
. On dual channel meters select the required channel.
6
Use the
and
keys to highlight the
Frequency
value field and press
to display the
Frequency
pop- up. Use the numeric keypad to
enter the required value in the
Frequency
pop- up window.
Figure 6-122
Frequency pop-up
7
Confirm your choice by pressing
or
.
8
Press
key to close the
Channel Setup
screen.
Now make the measurement.
9
Connect the power sensor to the signal to be measured.
The corrected measurement result is displayed.
N O T E
You can reduce the steps required to carry out the zero and calibration procedure as
follows:
1
Connect the power sensor to the POWER REF output.
2
Press
and
. (For dual channel meters, press
,
or
as required).
Zero + Cal
Zero + Cal
Zero + Cal A
Zero + Cal B
MHz
GHz