7
Using 8480 Series Power Sensors
186
N1911A/1912A P-Series Power Meters User’s Guide
5
To change the frequency, press
and use the
and
keys to
highlight the
Frequency
field.
6
Press
to display the
Frequency
pop- up window. Use the numeric
keypad to enter the required value in the
Frequency
pop- up window.
7
To confirm your choice, press the appropriate unit softkey.
8
Connect the power sensor to the signal to be measured.
9
The corrected measurement result is now displayed.
Figure 7-130
Frequency/calibration table display
N O T E
If the measurement frequency does not correspond directly to a frequency in the sensor
calibration table, the power meter calculates the calibration factor using linear
interpolation.
If you enter a frequency outside the frequency range defined in the sensor calibration table,
the power meter uses the highest or lowest frequency point in the sensor calibration table
to set the calibration factor.
N O T E
When
display mode is chosen, the frequency you entered and the
sensor table identifier is displayed in the upper window. Also, pressing
,
displays the frequency you entered and calibration factor for each channel derived from the
selected sensor tables.
Single Numeric
Offset