Using N8480 Series Power Sensors
198
N1911A/1912A P-Series Power Meters User’s Guide
Power Meter Configuration Changes
The Agilent P- Series power meter recognizes when an Agilent N8480
Series power sensor is connected. The N8480 Series power sensors
(excluding Option CFT) sensor calibration data is automatically read by
the power meter. In addition, the auto- averaging settings shown in
Figure 8- 134
are automatically configured.
Figure 8-134
Auto-averaging settings
Sensor Power
Minimum
Se
nsor Dynamic
Range
1
2
3
4
1
1
2
8
Resolution Setting
Number of A
verages
2
2
4
32
2
4
32
2
2
16
256
2
8
128
128
Maximum
4
64
256
512
U
pper Range
Lowe
r Range
Sensor Power
0 dBm
–1 dBm
–30 dBm
–10 dBm
–20 dBm
2
20 dBm
17 dBm
–10 dBm
10 dBm
0 dBm
N8481/2H
30 dBm
29 dBm
0 dBm
20 dBm
10 dBm
N8481/2B
N8481/2/5/7A
N8486AQ/AR
N O T E
These values are valid only for the power meter channel connected to an Agilent N8480
Series power sensor. Averaging settings can also be manually configured. Refer to
“Setting
Measurement Averaging”
on page 49 for more details.