User’s Guide
Agilent Technologies 34136A High Voltage Probe

W A R N I N G

This high voltage probe is designed to prevent accidental shock to the operator when properly used.  
This operating note must be read and understood prior to using the probe.  Improper procedures or 
incorrect analysis of the measurement situation can result in serious shock.

General Information

The Agilent Technologies 34136A High Voltage Probe is an accessory to be used with analog or digital 
multimeters having an input resistance of 10 M

 Ω ( ±1%) .

  The Agilent Technologies 34136A is a 

1000:1 divider which extends a voltmeter’s measurement capability to 40 kV dc/ 28 kV rms ac.

Specifications

Other Characteristics

AC output derates with frequency to 
-3db at 300 Hz. The graph at the right 
shows typical AC accuracy.

Actual AC accuracy varies with the 
amount of shunted input capacitance.

Maximum Input Voltage

Temperature Coefficient

DC Voltage:  40kV
AC Voltage:  (0 to 300 Hz) 28 kV 
RMS

Less than 200ppm/

 °

C

Accuracy

Division Ratio Accuracy

DC volts: 

 ±2% (1

 kV to 20 kV)

±3%  (20

 kV to 40 kV)

 

1000:1 ±1% 

when terminated in 10 M

Ω

2000:1  

±6% 

when terminated in 1 M

Ω

 

Note:

 For the 34401A, 34405A, L4411A, 34410A/11A, 

and 3458A multimeters, the ACV input impedance function 
has a 1 M

Ω

 input impedance. 

Pollution Degree 2

Input Resistance

Operating Temperature

1000 M

Ω

0  °

C to +50

 °

C

Cable Length:

Storage Temperature:

1 meter

−20 °

C to +70 

 °

C

Altitude:

Humidity

Up to 4,600 meters (15,000 ft)

Up to 80

relative humidity at +40 

 

°

C

Accuracy

AC Volts:  5% at 60 Hz

-3

0

Atten
(dB)

Frequency

DC

300 Hz