User’s Guide
Agilent Technologies 34136A High Voltage Probe
W A R N I N G
This high voltage probe is designed to prevent accidental shock to the operator when properly used.
This operating note must be read and understood prior to using the probe. Improper procedures or
incorrect analysis of the measurement situation can result in serious shock.
General Information
The Agilent Technologies 34136A High Voltage Probe is an accessory to be used with analog or digital
multimeters having an input resistance of 10 M
Ω ( ±1%) .
The Agilent Technologies 34136A is a
1000:1 divider which extends a voltmeter’s measurement capability to 40 kV dc/ 28 kV rms ac.
Specifications
Other Characteristics
AC output derates with frequency to
-3db at 300 Hz. The graph at the right
shows typical AC accuracy.
Actual AC accuracy varies with the
amount of shunted input capacitance.
Maximum Input Voltage
Temperature Coefficient
DC Voltage: 40kV
AC Voltage: (0 to 300 Hz) 28 kV
RMS
Less than 200ppm/
°
C
Accuracy
Division Ratio Accuracy
DC volts:
±2% (1
kV to 20 kV)
±3% (20
kV to 40 kV)
1000:1 ±1%
when terminated in 10 M
Ω
2000:1
±6%
when terminated in 1 M
Ω
Note:
For the 34401A, 34405A, L4411A, 34410A/11A,
and 3458A multimeters, the ACV input impedance function
has a 1 M
Ω
input impedance.
Pollution Degree 2
Input Resistance
Operating Temperature
1000 M
Ω
0 °
C to +50
°
C
Cable Length:
Storage Temperature:
1 meter
−20 °
C to +70
°
C
Altitude:
Humidity
Up to 4,600 meters (15,000 ft)
Up to 80
%
relative humidity at +40
°
C
Accuracy
AC Volts: 5% at 60 Hz
-3
0
Atten
(dB)
Frequency
DC
300 Hz