Features and Functions
2
34405A User’s and Service Guide
35
Math Operations and the Secondary Display
When a math operation is selected, the secondary display shows the result
of the math operation or the value(s) being used by the math operation.
For example, a typical primary display showing the Limit math operation
for DCV measurements and a secondary display showing a HI limit
exceeded is:
Repeatedly pressing
cycles through the secondary display choices
for the present math operation as shown in the table below. (
Reading
is
used in the table below to indicate the original measured reading value.)
mV DC
Limit
Disp
Secondary Display
Math Operation
Primary Display
Default Secondary
Display
Press Disp
Key Once
Press Disp
Key Twice
Press Disp
Key Three Times
Press Disp
Key Four Times
Null
Nulled Reading
Reference Value
Off
dBm
dBm
Present Reading
R
REF
Off
dB
dB
Present Reading
dB Ref (in
dBm)
Off
Min/Max
Reading
Max value
Min value
Avg value
N (count) value
Off
Limit
Reading
PASS
HI
LO
HI Limit
LO Limit
Off
Hold
Held Reading
Present Reading
Off
Downloaded from
www.Manualslib.com
manuals search engine