Features and Functions
2
34405A User’s and Service Guide
43
Reset/Power-On State
The table below summarizes the 34405A's settings as received from the
factory, following power cycling, and following the *RST command received
over the USB remote interface. Non- volatile, user customizable behavioral
differences are shown in
BOLD
type.
Table 2
Reset/Power-On State
Parameter
Factory Setting
Power-on / Reset State
Measurement Configuration
Function
DCV
DCV
Range
AUTO
AUTO
Resolution
5-½ digits
5-½ digits
Temperature Units
°C
User setting
Math Operations
Math State, Function
Off, Null
Off, Null
Math Registers
Cleared
Cleared
dBm Reference Resistance
600
Ω
User setting
Math Register Editing
On
User setting
Trigger Operations
Trigger Source*
Auto Trigger (Local Mode)
IMMediate (Remote Mode)
Auto Trigger (Local Mode)
IMMediate (Remote Mode)
System-Related Operations
Power-Down Recall
Disabled
User Setting
Stored States
0-4 cleared
No Change
Beeper
On
User Setting
Display
On
On
Remote/ Local State*
Local
Local
Keyboard*
Unlocked, Local key enabled
Unlocked, Local key enabled