74
34405A User’s and Service Guide
4
Performance Tests and Calibration
DC Current Gain Verification Test
1
Connect the calibrator to the front panel I and LO input connectors.
2
Select each function and range in the order shown below. Provide the
input shown in the table below.
3
Make a measurement and observe the result. Compare measurement
results to the appropriate test limits shown in the table. (Be certain to
allow for appropriate source settling when using the Fluke 5520A.)
[1] Select 5½- digit measurement resolution
Q
= Quick performance verification test points
Table 6
DC Current Gain Verification Test
Input
Function
[1]
Range
Quick Check
Error from Nominal
1 year
10mA
DC Current
10 mA
Q
± 6.5µA
100mA
100 mA
± 55µA
1A
1 A
Q
± 2.07mA
Caution: Connect calibrator to multimeter’s 12A and LO terminals before applying 10A
10A
10 A
± 25.7mA
Downloaded from
www.Manualslib.com
manuals search engine