Performance Tests and Calibration
4
34405A User’s and Service Guide
75
Ohms Gain Verification Test
Configuration: 2- Wire Ohms (CONFigure:RESistance)
1
Select the Ohms function.
2
Select each range in the order shown below. Provide the resistance
value indicated. Compare measurement results to the appropriate test
limits shown in the table. (Be certain to allow for appropriate source
settling.)
[1] Select 5½- digit measurement resolution
[[2] Specifications are for 2- wire ohms function using the Null math function
enabled to eliminate lead resistance. Without Null, add 0.2
Ω
additional error.
Q
= Quick performance verification test points
Table 7
Ohms Gain Verification Test
Input
Function
[1]
Range
Quick Check
Error from Nominal
1 year
100
Ω
2-Wire Ohms
100
Ω
±58 m
Ω
[2]
1 k
Ω
1 k
Ω
Q
±550 m
Ω
[2]
10 k
Ω
10 k
Ω
±5.6
Ω
[2]
100 k
Ω
100 k
Ω
±57
Ω
1 M
Ω
1 M
Ω
±670
Ω
10 M
Ω
10 M
Ω
Q
±25.5 k
Ω
100 M
Ω
100 M
Ω
±2.005M
Ω
Downloaded from
www.Manualslib.com