76
34405A User’s and Service Guide
4
Performance Tests and Calibration
Frequency Gain Verification Test
Configuration: Frequency (CONFigure:FREQuency)
1
Select the Frequency function.
2
Select each range in the order shown below. Provide the input voltage
and frequency indicated. Compare measurement results to the
appropriate test limits shown in the table. (Be certain to allow for
appropriate source settling.)
[1] Select 5½- digit measurement resolution
Q
= Quick performance verification test points
Table 8
Frequency Gain Verification Test
Voltage
Input
Frequency
Function
[1]
Range
Quick Check
Error from Nominal
1 year
200mVrms
1kHz
Frequency
1V
Q
±0.23Hz
200mVrms
10kHz
1V
±2.3Hz
Downloaded from
www.Manualslib.com