Performance Tests and Calibration
4
34405A User’s and Service Guide
77
AC Voltage Verification Test
Configuration: AC Volts (CONFigure[:VOLTage]:AC)
1
Select the AC Voltage function.
2
Select each range in the order shown below. Provide the indicated input
voltage and frequency. Compare measurement results to the appropriate
test limits shown in the table. (Be certain to allow for appropriate
source settling.)
[1] Select 5½- digit measurement resolution
Q
= Quick performance verification test points
Table 9
AC Volts Verification Test
Vrms
Input Frequency
Function
[1]
Range
Quick Check
Error from Nominal
1 year
100mV
1kHz
AC Voltage
100mV
± 0.3 mV
100mV
30kHz
100mV
± 1.8 mV
100mV
100kHz
100mV
± 5.3 mV
1V
1kHz
1V
Q
± 3.0 mV
1V
30kHz
1V
± 11 mV
1V
100kHz
1V
± 32 mV
10V
45Hz
10V
± 110 mV
10V
1kHz
10V
± 30 mV
10V
30kHz
10V
Q
± 0.11 V
10V
100kHz
10V
± 0.32 V
100V
1kHz
100V
Q
± 0.3 V
100V
30kHz
100V
± 1.1 V
100V
100kHz
100V
± 3.2 V
750V
1kHz
750V
± 2.25 V
Caution: Set the calibrator output to 0V before disconnecting it from the multimeter input terminals.