Performance Tests and Calibration
4
34405A User’s and Service Guide
79
Optional AC Voltage Performance Verification Test
Configuration: AC Volts (CONFigure[:VOLTage]:AC)
1
Select the AC Voltage function.
2
Select each range in the order shown below. Provide the indicated input
voltage and frequency. Compare measurement results to the appropriate
test limits shown in the table. (Be certain to allow for appropriate
source settling.)
[1] Select 5½- digit measurement resolution
Table 11
Optional AC Voltage Performance Verification Test
Vrms
Input Frequency Function
[1]
Range
Error from Nominal
1 year
1V
45Hz
AC Voltage
1V
±11mV
1V
1kHz
1V
±3mV
1V
10kHz
1V
±3mV
1V
30kHz
1V
±11mV
1V
100kHz
1V
±32mV
10V
1kHz
10V
±30mV
1V
1kHz
10V
±12mV
0.1V
1kHz
10V
±10.2mV
Downloaded from