124
34410A/11A User’s Guide
5
Specifications
Performance Versus Integration Time – 60Hz (50Hz) Power line frequency
[ 1 ] Resolution is defined as the typical 10 VDC range RMS noise.
[ 2 ] Normal mode rejection for power–line frequency ± 0.1%.
[ 3 ] Maximum rate for DCV, DCI, and 2–Wire resistance functions
(using zero settling delay, autozero off, etc.).
[ 4 ] Autozero on for => 1 NPLC.
[ 5 ] For power–line frequency ± 1% 75 dB and for ± 3% 55 dB.
[ 6 ] Only for 34411A.
Transfer Accuracy (Typical)
All DC volts, <0.12 A DC Current, < 1.2 M
W
: (24 hour % of range error) / 2)
All other DC current and resistance: (24 hour % of range error + % of reading)/2
Conditions:
- Within 10 minutes and ±0.5 °C
- Within ±10% of initial value.
- Following a 2–hour warm–up.
- Fixed range.
- Using >= 10.
- Measurements are made using accepted metrology practices.
RMS Noise adder % range
[4]
Integration Time
Number of Power
Line Cycles (NPLC)
Resolution
ppm Range
[1]
NMR db
[2]
Readings /
Second
[3]
DCV
10, 1000 V
DCV
1, 100 V
Resistance
1K, 10K ohm
DCV
0.1 V
Resistance
100 ohm
DCI
1 amp
0.001
[6]
0.002
[6]
0.006
0.02
0.06
0.2
1
2
10
100
30
15
6
3
1.5
0.7
0.3
0.2
0.1
0.03
0
0
0
0
0
0
55
110 [
5]
110
[5]
110
[5]
50,000
25,000
10,000
3000
1000
300
60(50)
30(25)
6(5)
0.6(0.5)
0.0060
0.0030
0.0012
0.0006
0.0003
0.0002
0.0
0.0
0.0
0.0
0.0100
0.0060
0.0040
0.0030
0.0020
0.0015
0.0001
0.0001
0.0
0.0
0.1000
0.0600
0.0600
0.0300
0.0200
0.0150
0.0010
0.0010
0.0005
0.0
Summary of Contents for 34410A
Page 1: ...Agilent Technologies Agilent 34410A 11A 6 Digit Multimeter User s Guide...
Page 6: ...6 34410A 11A User s Guide...
Page 12: ...12 34410A 11A User s Guide...
Page 34: ...34 34410A 11A User s Guide 1 Quick Start...
Page 96: ...96 34410A 11A User s Guide 3 Remote Interface Configuration...
Page 106: ...106 34410A 11A User s Guide 4 Measurement Tutorial...
Page 120: ...120 34410A 11A User s Guide 4 Measurement Tutorial...
Page 140: ...140 34410A 11A User s Guide 5 Specifications...