34410A/11A User’s Guide
125
Specifications
5
DC Voltage
Measurement Method:
Continuously integrating multi–slope IV
10 VDC Linearity:
0.0002% of r 0.0001% of range
Input Resistance:
0.1 V, 1 V, 10 V Ranges
Selectable 10 M
W
or >10 G
W
(For these ranges, inputs beyond ±17 V are clamped through 100 k
W
typical)
100 V, 1000 V Ranges
10 M
W
±1%
Input Bias Current:
< 30 pA at 25 °C
Input Terminals:
Copper alloy
Input Protection:
1000 V
DC CMRR
140 dB for 1 k
W
unbalance in LO lead. ±500 VDC maximum
Resistance
Measurement Method:
Selectable 4–wire or 2–wire ohms.
Max. Lead Resistance
10% of range per lead for 100
W
, 1 k
W
ranges.
(4–wire ohms)
1 k
W
per lead on all other ranges
Input Protection:
1000 V on all ranges
Offset Compensation:
Selectable on the 100
W
, 1 k
W
, and 10 k
W
ranges
DC Current
Shunt Resistor:
0.1
W
for 1 A, 3 A.
2
W
for 10 mA, 100 mA.
200
W
for 100 µA, 1 mA.
Input Protection:
Externally accessible 3 A, 250 V fuse
Continuity / Diode Test
Response Time:
300 samples / sec with audible tone
Continuity Threshold:
Fixed at 10
W
Autozero OFF Operation (Typical)
Following instrument warm–up at a stable ambient temperature ±1 °C and <5 minutes.
Add 0.0002% of range + 2 µV for DCV or + 2 m
W
for resistance.
Settling Considerations
Reading settling times are affected by source impedance, cable dielectric characteristics, and input
signal changes. Default delays are selected to give first reading right for most measurements.
Measurement Considerations
Agilent recommends the use of Teflon or other high–impedance, low–dielectric absorption wire
insulation for these measurements.
Summary of Contents for 34410A
Page 1: ...Agilent Technologies Agilent 34410A 11A 6 Digit Multimeter User s Guide...
Page 6: ...6 34410A 11A User s Guide...
Page 12: ...12 34410A 11A User s Guide...
Page 34: ...34 34410A 11A User s Guide 1 Quick Start...
Page 96: ...96 34410A 11A User s Guide 3 Remote Interface Configuration...
Page 106: ...106 34410A 11A User s Guide 4 Measurement Tutorial...
Page 120: ...120 34410A 11A User s Guide 4 Measurement Tutorial...
Page 140: ...140 34410A 11A User s Guide 5 Specifications...