34410A/11A User’s Guide
131
Specifications
5
Measurement and System Speeds
DMM Measurements Speeds
[1] ½ scale input signal, immediate trigger, trigger delay 0, autozero off, autorange off, no math,
60 Hz line, null off, sample count 1, trig count 1
Direct I/O Measurements
[1]
Single Reading – Measure and I/O Time
Measurement
Into Memory
(Readins/Sec)
(VM Complete)
Function
Resolution
(NPLC)
GPIB
Sec
USB 2.0
Sec
LAN (VXI-11)
Sec
LAN (Sockets)
Sec
DCV (10 V Range) 0.006
0.0030 0.0036 0.0050 0.0035
10000
0.06
0.0035 0.0044 0.0048 0.0043
1000
1
0.0200
0.0200
0.0200
0.0200
60
ACV (10 V Range)
Slow Filter
0.0100
0.0100
0.0100
0.0100
50
Medium Filter
0.0100
0.0100
0.0100
0.0100
150
Fast Filter
0.0100
0.0100
0.0100
0.0100
500
2–Wire
W
(10 k
W
Range)
0.006
0.0035
0.0036
0.0049
0.0035
10000
0.06
0.0043
0.0045
0.0049
0.0043
1000
1
0.0195
0.0200
0.0200
0.0200
60
4–wire
W
(10 k
W
Range)
0.006
0.0054
0.0040
0.0045
0.0056
1200
0.06
0.0074
0.0078
0.0078
0.0074
380
1
0.0390
0.0390
0.0390
0.0390
30
Frequency
1 KHz, 10 V Range
Fast Filter
1 ms Gate
0.0100
0.0100
0.0100
0.0100
500
10 mS Gate
0.0200
0.0200
0.0200
0.0200
80
100 mS Gate
0.1150
0.1150
0.1130
0.1130
10
1 S Gate
1.0200 1.0200 1.0200 1.0200
1
Capacitance (100 nF Rage)
0.0820
0.0820
0.0820
0.0820
11
Summary of Contents for 34410A
Page 1: ...Agilent Technologies Agilent 34410A 11A 6 Digit Multimeter User s Guide...
Page 6: ...6 34410A 11A User s Guide...
Page 12: ...12 34410A 11A User s Guide...
Page 34: ...34 34410A 11A User s Guide 1 Quick Start...
Page 96: ...96 34410A 11A User s Guide 3 Remote Interface Configuration...
Page 106: ...106 34410A 11A User s Guide 4 Measurement Tutorial...
Page 120: ...120 34410A 11A User s Guide 4 Measurement Tutorial...
Page 140: ...140 34410A 11A User s Guide 5 Specifications...