34410A/11A User’s Guide
25
Quick Start
1
To Make a 2-Wire Resistance Measurement
•
Ranges: 100
W
, 1 k
W
, 10 k
W
, 100 k
W
, 1 M
W
, 10 M
W
, 100 M
W
•
Maximum resolution: 100
mW
(
on 100
W
range
)
•
Configurable parameters:
INTEGRATION
,
RANGE
,
OFFSET COMP
,
AUTO ZERO
,
NULL
,
and
NULL VALUE
To null
–
out the test lead resistance:
1
Connect one end of the test leads at the meter, and short the probe
ends together.
2
Press null.
3
Connect the probe ends to the test circuit, and measure the corrected
resistance value.
To Make a 4-wire Resistance Measurement
•
Ranges: 100
W
, 1 k
W
, 10 k
W
, 100 k
W
, 1 M
W
, 10 M
W
, 100 M
W
•
Maximum resolution: 100
mW
(
on 100
W
range
)
•
Configurable parameters:
INTEGRATION
,
RANGE
,
OFFSET COMP
,
AUTO ZERO
,
NULL
,
and
NULL VALUE