4
34410A/11A User’s Guide
Protection Limits
The Agilent 34410A Digital Multimeter pro-
vides protection circuitry to prevent damage
to the instrument and to protect against the
danger of electric shock, provided the Pro-
tection Limits are not exceeded. To ensure
safe operation of the instrument, do not
exceed the Protection Limits shown on the
front and rear panel, and defined below:
Note: The front-panel terminals are shown
above. The rear-panel terminals are identi-
cal. The Front/Rear switch selects the ter-
minal set to be used. Do not operate this
switch while signals are present on the
front or rear terminals. The current-protec-
tion fuse is on the rear panel.
Input Terminal Protection
Limits
Protection Limits are defined for the three
input terminals:
Main Input (HI and LO) Terminals. The HI
and LO input terminals are used for voltage,
resistance, capacitance, and diode test
measurements. Two Protection Limits are
defined for these terminals:
HI to LO Protection Limit. The Protection
Limit from HI to LO ("A" in the figure at
left) is 1000 VDC or 750 VAC, which is
also the maximum voltage measurement.
This limit can also be expressed as 1000
Vpk maximum.
LO to Ground Protection Limit. The LO
input terminal can safely "float" a maxi-
mum of 500 Vpk relative to ground. This is
Protection Limit "B" in the figure.
As is implied by the above limits, the Protec-
tion Limit for the HI input terminal is a maxi-
mum of 1500 Vpk relative to ground.
Current Input Terminal. The current input
("I") terminal has a Protection Limit of 3A
(rms) maximum current flowing from the LO
input terminal. This is Protection Limit "C"
in the figure. Note that the current input ter-
minal will be at approximately the same
voltage as the LO terminal.
Note: The current-protection circuitry
includes a fuse on the rear panel. To main-
tain protection, replace this fuse only with a
fuse of the specified type and rating (see
"Operator Maintenance" in Chapter 1).
Sense Terminal Protection
Limits
The HI and LO sense terminals are used
only for four-wire resistance measurements
("
Ω
4W"). The Protection Limit is 200 Vpk
for all of the terminal pairings ("D" in the
figure):
LO sense to LO input.
HI sense to LO input.
HI sense to LO sense.
Note: The 200 Vpk limit on the sense termi-
nals is the Protection Limit. Operational
voltages in resistance measurements are
much lower — less than 10 V in normal
operation.
IEC Measurement Category II
Overvoltage Protection
To protect against the danger of electric
shock, the Agilent 34410A/11A Digital Mul-
timeter provides overvoltage protection for
line-voltage mains connections meeting
both of the following conditions:
The HI and LO input terminals are con-
nected to the mains under Measurement
Category II conditions, defined below, and
The mains are limited to a maximum line
voltage of 300 VAC.
IEC Measurement Category II includes elec-
trical devices connected to mains at an out-
let on a branch circuit. Such devices include
most small appliances, test equipment, and
other devices that plug into a branch outlet
or socket. The 34410A/11A may be used to
make measurements with the HI and LO
inputs connected to mains in such devices,
or to the branch outlet itself (up to 300
VAC). However, the 34410A/11A may not be
used with its HI and LO inputs connected to
mains in permanently installed electrical
devices such as the main circuit-breaker
panel, sub-panel disconnect boxes, or per-
manently wired motors. Such devices and
circuits are subject to overvoltages that may
exceed the protection limits of the 34410A.
Note: Voltages above 300 VAC may be mea-
sured only in circuits that are isolated from
mains. However, transient overvoltages are
also present on circuits that are isolated
from mains. The Agilent 34410A/11A is
designed to safely withstand occasional
transient overvoltages up to 2500 Vpk. Do
not use this equipment to measure circuits
where transient overvoltages could exceed
this level.
Summary of Contents for 34410A
Page 1: ...Agilent Technologies Agilent 34410A 11A 6 Digit Multimeter User s Guide...
Page 6: ...6 34410A 11A User s Guide...
Page 12: ...12 34410A 11A User s Guide...
Page 34: ...34 34410A 11A User s Guide 1 Quick Start...
Page 96: ...96 34410A 11A User s Guide 3 Remote Interface Configuration...
Page 106: ...106 34410A 11A User s Guide 4 Measurement Tutorial...
Page 120: ...120 34410A 11A User s Guide 4 Measurement Tutorial...
Page 140: ...140 34410A 11A User s Guide 5 Specifications...