50
34410A/11A User’s Guide
2
Features and Functions
During a measurement session, configuration settings may be saved in one
of four Stored States, for recall after a power–on or reset operation.
See
for more information.
System
–
Related Operations
Factory Setting
●
Beeper Mode
●
On
●
Radix Character
●
Period
●
Thousands Seperator
●
On
Display State
On
Reading Memory
Cleared
Error Queue
Cleared at Power
–
On; not
cleared
by a front panel or remote interface
Reset command
●
Stored States
●
No change
●
Calibration State
●
Secured
Input/Output Configuration
Factory Setting
Enable Interfaces:
*
●
LAN
●
Enabled
●
GPIB
●
Enabled
●
USB
●
Enabled
Lan Settings:
*
●
DHCP
●
On
●
Auto IP
●
On
●
IP Address
●
169.254.4.10
**
●
Subnet Mask
●
255.255.0.0
●
Default Gateway
●
0.0.0.0
●
DNS Server
●
0.0.0.0
●
Host Name
●
A-34410A-nnnn
**
●
LAN Services
●
Enable All
GPIB Settings:
*
●
GPIB Address
●
22
*
Interface enable or LAN setting changes
take effect only after a CPU reboot
**
For model 34411A, the default
IP address is 169.254.4.11 and
host name is A-34411A-nnnn;
also note that nnnn is the last 4
digits of the serial number.
Summary of Contents for 34410A
Page 1: ...Agilent Technologies Agilent 34410A 11A 6 Digit Multimeter User s Guide...
Page 6: ...6 34410A 11A User s Guide...
Page 12: ...12 34410A 11A User s Guide...
Page 34: ...34 34410A 11A User s Guide 1 Quick Start...
Page 96: ...96 34410A 11A User s Guide 3 Remote Interface Configuration...
Page 106: ...106 34410A 11A User s Guide 4 Measurement Tutorial...
Page 120: ...120 34410A 11A User s Guide 4 Measurement Tutorial...
Page 140: ...140 34410A 11A User s Guide 5 Specifications...