60
34410A/11A User’s Guide
2
Features and Functions
•
Remote Interface Operation:
For all measurements except temperature,
diode check and continuity:
Autoranging may be enabled or disabled for the specified <function>
using this SENSe command:
SENSe:<function>:RANGe:AUTO {OFF
>
ONCE
>
0
>
ON
>
1};
where
<function>
=
VOLTage:DC
,
VOLTage:AC
,
CURRent:DC
,
CURRent:AC,RESistance
,
FRESistance
or
FREQency
The following command queries the autoranging feature status for the
specified <function>:
SENSe:<function>:RANGe:AUTO?
This query command returns “0” (OFF) or “1” (ON).
The following command manually sets the multimeter range for the
specified <function>:
SENSe:<function>:RANGe[:UPPER] {<range>
>
MIN
>
MAX
>
DEF}
Note:
for
<function>
=
FREQuency
, this command changes the ac
input range.
A manual range may also be set using the
CONFigure
or
MEASure
commands.
The following command queries the multimeter range setting for the
specified <function>:
SENSe:<function>:RANGe[:UPPER]? [{MIN
>
MAX}]
Refer to the
Agilent 34410A/34411A Programmer’s Reference
for
complete description and syntax for all commands on this page.
Summary of Contents for 34410A
Page 1: ...Agilent Technologies Agilent 34410A 11A 6 Digit Multimeter User s Guide...
Page 6: ...6 34410A 11A User s Guide...
Page 12: ...12 34410A 11A User s Guide...
Page 34: ...34 34410A 11A User s Guide 1 Quick Start...
Page 96: ...96 34410A 11A User s Guide 3 Remote Interface Configuration...
Page 106: ...106 34410A 11A User s Guide 4 Measurement Tutorial...
Page 120: ...120 34410A 11A User s Guide 4 Measurement Tutorial...
Page 140: ...140 34410A 11A User s Guide 5 Specifications...