34410A/11A User’s Guide
63
Features and Functions
2
Beeper
Normally, the multimeter will emit a tone from the front panel whenever
certain conditions are met (e.g. the multimeter will beep when a stable
reading is captured in reading hold mode). The beeper is factory set to
ON
,
but may be disabled or enabled manually.
Only when the beeper is
enabled
, the multimeter will emit a tone when:
1
A set
LO
or
HI
limit is exceeded in a limit test.
2
A stable reading is captured in reading hold mode.
3
A forward–biased diode is measured in the diode test function.
Disabling
the beeper will cancel the tone in the three cases listed above.
However, the multimeter will
always
emit a tone when:
1
An error is generated.
2
The continuity threshold is exceeded.
3
A
SYST:BEEP
command is sent.
•
Front Panel Operation:
Press
to display
UTILITY MENU
,
and
press
as needed until
MISC SETTINGS
is displayed. Press
as
needed
to
display
BEEPER
,
press
or
to select
OFF
or
ON
, and press
. Then step through or exit the configuration menu.
•
Remote Interface Operation:
The following command disables or enables the
front panel beeper:
SYSTem:BEEPer:STATe {OFF
>
0
>
ON
>
1}
The following command queries the state of the beeper:
SYSTem:BEEPer:STATe?
This query returns a “0” (OFF) or “1” (ON).
Refer to the
Agilent 34410A/34411A Programmer’s Reference
for
complete description and syntax for these commands.
Summary of Contents for 34410A
Page 1: ...Agilent Technologies Agilent 34410A 11A 6 Digit Multimeter User s Guide...
Page 6: ...6 34410A 11A User s Guide...
Page 12: ...12 34410A 11A User s Guide...
Page 34: ...34 34410A 11A User s Guide 1 Quick Start...
Page 96: ...96 34410A 11A User s Guide 3 Remote Interface Configuration...
Page 106: ...106 34410A 11A User s Guide 4 Measurement Tutorial...
Page 120: ...120 34410A 11A User s Guide 4 Measurement Tutorial...
Page 140: ...140 34410A 11A User s Guide 5 Specifications...