34410A/11A User’s Guide
71
Features and Functions
2
•
The configured range mode (auto or manual) will be used for reading
hold.
•
To enable reading hold
, press
to display
TRIGGER
.
Press
as needed to select
HOLD
, and press
.
Internal Triggering
The internal triggering mode is available from the remote interface only.
In the internal trigger mode, the trigger signal is always present. When
you place the multimeter in the wait–for–trigger state, the trigger is issued
immediately. This is the default trigger source for remote interface
operation.
•
Remote Interface Operation:
The following command selects the internal
trigger source:
TRIGger:SOURce IMMediate
The
CONFigure
and
MEASure?
commands automatically set the trigger
source to
IMMediate
.
Refer to the
Agilent 34410A/34411A Programmer’s Reference
for
complete description and syntax for these commands.
Software (Bus ) Triggering
The bus trigger mode is available from the remote interface only.
The bus trigger mode corresponds in function to pressing the
key
from the front panel, except that it is initiated by sending a bus trigger
command, after selecting BUS as the trigger source.
•
The following comand selects the bus trigger source:
TRIGger:SOURce BUS
Then, any of the following commands will initiate readings:
MEASure?
READ?
INITiate
Refer to the
Agilent 34410A/34411A Programmer’s Reference
for
complete description and syntax for these commands.
Summary of Contents for 34410A
Page 1: ...Agilent Technologies Agilent 34410A 11A 6 Digit Multimeter User s Guide...
Page 6: ...6 34410A 11A User s Guide...
Page 12: ...12 34410A 11A User s Guide...
Page 34: ...34 34410A 11A User s Guide 1 Quick Start...
Page 96: ...96 34410A 11A User s Guide 3 Remote Interface Configuration...
Page 106: ...106 34410A 11A User s Guide 4 Measurement Tutorial...
Page 120: ...120 34410A 11A User s Guide 4 Measurement Tutorial...
Page 140: ...140 34410A 11A User s Guide 5 Specifications...