34410A/11A User’s Guide
73
Features and Functions
2
•
Front Panel Operation:
To manually specify the trigger delay, press
to display
TRIGGER
. Press
twice to select
ADVANCED
,
and press
as needed to display
TRIG DELAY
. Press
as needed to
select
MANUAL
, and press
.
Use the navigation keypad to set the
desired trigger delay, and press
. Then step through or exit the
advanced triggering menu.
To specify a trigger delay of
zero
, press
to display
TRIGGER
. Press
twice to select
ADVANCED
, and press
as needed to
display
TRIG DELAY
. Press
or
as needed to select
ZERO
, and press
.
Then step through or exit the advanced triggering menu.
•
Remote Interface Operation:
The following command specifies the trigger
delay, in seconds:
TRIGger:DELay {<seconds>
>
MIN
>
MAX}
If the multimeter is configured to take more than one sample per
trigger, the trigger delay you specify is inserted between the trigger and
each
reading (if the sample source is set to
AUTO
), or only before the
first reading (if the sample source is set to
TIMer
). See the
SAMPle:SOURce
command description for more information.
After setting the trigger delay, you must use the
INITiate
or
READ
command to place the meter in the wait–for–trigger state before
readings are triggered
(see
.
The
MEASure?
command sets the trigger delay to AUTO.
Refer to the
Agilent 34410A/34411A Programmer’s Reference
for
complete description and syntax for these command.
Summary of Contents for 34410A
Page 1: ...Agilent Technologies Agilent 34410A 11A 6 Digit Multimeter User s Guide...
Page 6: ...6 34410A 11A User s Guide...
Page 12: ...12 34410A 11A User s Guide...
Page 34: ...34 34410A 11A User s Guide 1 Quick Start...
Page 96: ...96 34410A 11A User s Guide 3 Remote Interface Configuration...
Page 106: ...106 34410A 11A User s Guide 4 Measurement Tutorial...
Page 120: ...120 34410A 11A User s Guide 4 Measurement Tutorial...
Page 140: ...140 34410A 11A User s Guide 5 Specifications...