74
34410A/11A User’s Guide
2
Features and Functions
Automatic Trigger Delay
If you do not specify a trigger delay manually, the multimeter selects the
trigger delay time automatically. Automatic trigger delays are set to ensure
that completely settled and accurate measurements are taken. The delay is
determined by function, range, integration time and ac filter setting, as
shown in the following tables
DC Voltage Measurements (all ranges):
DC Current Measurements (all ranges):
2
–
wire Resistance Measurements :
4
–
wire Resistance Measurements
Integration Time
Trigger Delay
¡
1 PLC
160
m
s
0.2 PLC, 0.06 PLC
130
m
s
~
0.02 PLC
100
m
s
Integration Time
Trigger Delay
¡
1 PLC
1.5 ms
~
0.2 PLC
1.0 ms
Range
Trigger Delay
(for
~
0.02 PLC)
Trigger Delay
(for 0.06 or 0.2 PLC)
Trigger Delay
(for
¡
1 PLC)
100
W
80
m
s
100
m
s
130
m
s
1
k
W
110
m
s
130
m
s
160
m
s
10 k
W
130
m
s
160
m
s
190
m
s
100 k
W
540
m
s
670
m
s
800
m
s
1 M
W
5.0 ms
6.0 ms
7.5 ms
10 M
W
–
1G
W
60 ms
70 ms
84 ms
Range
Trigger Delay
(for
~
0.02 PLC)
Trigger Delay
(for 0.06 or 0.2 PLC)
Trigger Delay
(for
¡
1 PLC)
100
W
–
100 k
W
1.0 ms
1.0 ms
1.5 ms
1
M
W
10 ms
10 ms
15 ms
10 M
W
–
1 G
W
100 ms
100 ms
100 ms
Summary of Contents for 34410A
Page 1: ...Agilent Technologies Agilent 34410A 11A 6 Digit Multimeter User s Guide...
Page 6: ...6 34410A 11A User s Guide...
Page 12: ...12 34410A 11A User s Guide...
Page 34: ...34 34410A 11A User s Guide 1 Quick Start...
Page 96: ...96 34410A 11A User s Guide 3 Remote Interface Configuration...
Page 106: ...106 34410A 11A User s Guide 4 Measurement Tutorial...
Page 120: ...120 34410A 11A User s Guide 4 Measurement Tutorial...
Page 140: ...140 34410A 11A User s Guide 5 Specifications...