78
34410A/11A User’s Guide
2
Features and Functions
Reading Memory
This feature is available from the remote interface only.
The multimeter’s reading memory is a first–in, first out (FIFO) buffer
holding up to 50,000 readings (model 34410A) or 1 million readings
(model 34411A). The oldest readings are preserved.
•
Remote Interface Operation:
The following command transfers readings
stored in non–volatile memory into the multimeter’s output buffer, from
where you can read them into your computer:
FETCh?
The following command will obtain whatever readings are present in
NVMEM, and erase them as they are read:
R?
Refer to the
Agilent 34410A/34411A Programmer’s Reference
for
complete description and syntax for these commands.