80
34410A/11A User’s Guide
2
Features and Functions
•
After the programmed count has been reached, the display will show
SAVING READINGS
, then
LOGGING COMPLETE
. The readings are stored in
non–volatile memory.
To stop a data logging event in progress, you may press either
,
or
. The primary display line will show
DATA LOGGER?
and
offers selections of
CONTINUE
or
STOP
. The data logger continues to take
readings while this is displayed.
•
If you select
STOP
and press
, the primary display will read
DATA
TO NV?
, prompted you to
SAVE
or
DISCARD
the readings already logged.
Select one of these options and press
to terminate the data
logging function.
•
If you select
CONTINUE
and press
, the data logging function will
continue as programmed, until complete.
To read the data logged from the front panel, press
to display
DATA LOGGER
. Press
to select
VIEW
–
RDGS,
and press
. Press
or
to scroll through the stored readings.
•
Remote Interface Operation:
The following command returns all readings
stored in non–volatile memory (NVMEM):
DATA:DATA? NVMEM
The following command deletes all readings stored in NVMEM.
DATA:DELete NVMEM
The following command returns the number of data points stored in
NVMEM
DATA:POINts? NVMEM
Refer to the
Agilent 34410A/34411A Programmer’s Reference
for
complete description and syntax for these commands.
Summary of Contents for 34410A
Page 1: ...Agilent Technologies Agilent 34410A 11A 6 Digit Multimeter User s Guide...
Page 6: ...6 34410A 11A User s Guide...
Page 12: ...12 34410A 11A User s Guide...
Page 34: ...34 34410A 11A User s Guide 1 Quick Start...
Page 96: ...96 34410A 11A User s Guide 3 Remote Interface Configuration...
Page 106: ...106 34410A 11A User s Guide 4 Measurement Tutorial...
Page 120: ...120 34410A 11A User s Guide 4 Measurement Tutorial...
Page 140: ...140 34410A 11A User s Guide 5 Specifications...