34410A/11A User’s Guide
123
Specifications
5
DC Characteristics
Accuracy Specifications ( % of r % of range )
[1]
[ 1 ] Specifications are for 1–hour warm–up and integration setting of 100 NPLC.
[ 2 ] Relative to calibration standards.
[ 3 ] 20% overrange on all ranges, except 1000 Vdc, 3 A range.
[ 4 ] Specifications are for 4–wire ohms function, or 2–wire ohms using Math Null. Without Math Null,
add 0.2
W
additional error in 2–wire ohms function.
[ 5 ] For each additional volt over ± 500 VDC add 0.02 mV of error.
[ 6 ] Accuracy specifications are for the voltage measured at the input terminals only. 1 mA test current
is typical. Variation in the current source will create some variation in the voltage drop across a
diode junction
Function
Range
[3]
Test Current or
Burden Voltage
24 Hour
[2]
T
CAL
± 1 °C
90 Day
T
CAL
± 5 °C
1 Year
T
CAL
± 5 °C
Temperature
Coefficient
0 °C – (T
CAL
– 5 °C)
(T
CAL
+ 5 °C) – 55 °C
DC
100.0000 mV
1.000000 V
10.00000 V
100.0000 V
1000.000 V [5]
0.0030+0.0030
0.0020+0.0006
0.0015+0.0004
0.0020+0.0006
0.0020+0.0006
0.0040+0.0035
0.0030+0.0007
0.0020+0.0005
0.0035+0.0006
0.0035+0.0006
0.0050+0.0035
0.0035+0.0007
0.0030+0.0005
0.0040+0.0006
0.0040+0.0006
0.0005+0.0005
0.0005+0.0001
0.0005+0.0001
0.0005+0.0001
0.0005+0.0001
Resistance
[4]
100.0000
W
1.000000 K
W
10.00000 K
W
100.0000 K
W
1.000000 M
W
10.00000 M
W
100.0000 M
W
1000.000 M
W
1 mA Current Source
1 mA
100 µA
10 µA
5.0 µA
500 nA
500 nA || 10 M
W
500 nA || 10 M
W
0.0030+0.0030
0.0020+0.0005
0.0020+0.0005
0.0020+0.0005
0.0020+0.0010
0.0100+0.0010
0.200+0.001
2.000+0.001
0.008+0.004
0.007+0.001
0.007+0.001
0.007+0.001
0.010+0.001
0.030+0.001
0.600+0.001
6.000+0.001
0.010+0.004
0.010+0.001
0.010+0.001
0.010+0.001
0.012+0.001
0.040+0.001
0.800+0.011
8.000+0.001
0.0006+0.0005
0.0006+0.0001
0.0006+0.0001
0.0006+0.0001
0.0010+0.0002
0.0030+0.0004
0.1000+0.0001
1.0000+0.0001
DC Current
100.0000 µA
1.000000 mA
10.00000 mA
100.0000 mA
1.000000 A
3.00000 A
<0.03 V Burden V
<0.3 V
<0.03 V
<0.3 V
<0.80 V
<2.0 V
0.010+0.020
0.007+0.006
0.007+0.020
0.010+0.004
0.050+0.006
0.100+0.020
0.040+0.025
0.030+0.006
0.030+0.020
0.030+0.005
0.080+0.010
0.120+0.020
0.050+0.025
0.050+0.005
0.050+0.020
0.050+0.005
0.100+0.010
0.150+0.020
0.0020+0.0030
0.0020+0.0005
0.0020+0.0020
0.0020+0.0005
0.0050+0.0010
0.0050+0.0020
Continunity
1000 Ohms
1 mA Test Current
0.002+0.010
0.008+0.020
0.010+0.020
0.0010+0.0020
Diode Test
1.0000 V
[6]
1 mA Test Current
0.002+0.010
0.008+0.020
0.010+0.020
0.0010+0.0020
Summary of Contents for 34410A
Page 1: ...Agilent Technologies Agilent 34410A 11A 6 Digit Multimeter User s Guide...
Page 6: ...6 34410A 11A User s Guide...
Page 12: ...12 34410A 11A User s Guide...
Page 34: ...34 34410A 11A User s Guide 1 Quick Start...
Page 96: ...96 34410A 11A User s Guide 3 Remote Interface Configuration...
Page 106: ...106 34410A 11A User s Guide 4 Measurement Tutorial...
Page 120: ...120 34410A 11A User s Guide 4 Measurement Tutorial...
Page 140: ...140 34410A 11A User s Guide 5 Specifications...