1
To Measure Capacitance
•
Ranges: 1 nF, 10 nF, 100 nF, 1
m
F, 10
m
F
•
Maximum resolution: 100 pF (
on 1 nF range
)
•
Configurable parameters:
RANGE
,
NULL
, and
NULL VALUE
To null
–
out the test lead capacitance:
1
Disconnect the + lead’s probe end from the test circuit, and leave open.
2
Press null.
3
Reconnect the + lead’s probe end to the test circuit, and measure the
corrected capacitance value.