34410A/11A User’s Guide
55
Features and Functions
2
•
Remote Interface Operation:
The following commands set the aperture
value, in seconds:
[SENSe:]VOLTage[:DC]:APERture {<seconds>
>
MIN
>
MAX
>
DEF}
[SENSe:]CURRent[:DC]:APERture {<seconds>
>
MIN
>
MAX
>
DEF}
[SENSe:]RESistance:APERture {<seconds>
>
MIN
>
MAX
>
DEF}
[SENSe:]FRESistance:APERture {<seconds>
>
MIN
>
MAX
>
DEF}
[SENSe:]TEMPerature:APERture {<seconds>
>
MIN
>
MAX
>
DEF}
The following command (similar queries for current, resistance and
temperature) queries the aperture setting:
[SENSe:]VOLTage[:DC]:APERture?[{MIN
>
MAX}]
The following command (similar queries for current, resistance and
temperature) queries if aperture mode is enabled:
[SENSe:]VOLTage[:DC]:APERture:ENABled?
This query returns a “0” (disabled) or “1” (enabled)
Refer to the
Agilent 34410A/34411A Programmer’s Reference
for
complete description and syntax for these commands.
•
The following table indicates the resulting resolution for each
integration time value (in power–line cycles) selected.
AC Measurements
See
on page 56 for configuration options for ac voltage, ac
current and frequency/period measurements.
Integration Time
Resolution
0.001 PLC
*
30 ppm x Range
0.002 PLC
*
15 ppm x Range
0.006 PLC
6.0 ppm x Range
0.02 PLC
3.0 ppm x Range
0.06 PLC
1.5 ppm x Range
0.2 PLC
0.7 ppm x Range
1 PLC (default)
0.3 ppm x Range
2 PLC
0.2 ppm x Range
10 PLC
0.1 ppm x Range
100 PLC
0.03 ppm x Range
*
These two settings are
provided on model 34411A only
.
Summary of Contents for 34410A
Page 1: ...Agilent Technologies Agilent 34410A 11A 6 Digit Multimeter User s Guide...
Page 6: ...6 34410A 11A User s Guide...
Page 12: ...12 34410A 11A User s Guide...
Page 34: ...34 34410A 11A User s Guide 1 Quick Start...
Page 96: ...96 34410A 11A User s Guide 3 Remote Interface Configuration...
Page 106: ...106 34410A 11A User s Guide 4 Measurement Tutorial...
Page 120: ...120 34410A 11A User s Guide 4 Measurement Tutorial...
Page 140: ...140 34410A 11A User s Guide 5 Specifications...