Chart Output (Analog Output)
You can use the Chart Output (Analog Output) connector on the rear
panel to run a strip chart recorder or similar instrument. The connector
provides an output voltage proportional to the measured voltage.
•
The output can range from –3.00 V to +3.00 V.
•
When chart output is disabled (Off), the output is set to 0 V.
•
The output impedance is 1 k
Ω
.
•
The output voltage is related to the input signal as follows:
(
reading
−
chart
null
)
span
=
chart
output
voltage
Note that
reading
includes the results of measurement unit
conversion, filtering, null, and any math operation performed on the
measurement.
•
The chart output state (including the
span
and
null
values) is stored
in
non-volatile
memory.
•
span
cannot have a negative value.
•
The chart output has a voltage "rollover" to prevent the loss of data,
Examples:
See page 38 for a menu procedure to enter a number.
1) To set the chart span so that a
±
1 mV input signal produces a
±
1.0
Volt chart output:
(
0.001
−
0.0
)
span
=
1.00
span
=
0.001
1.00
=
0.001
When chart null is set to 0, the value
1
span
can be considered the
gain
and
chart
output
=
gain
∗
reading
.
Chapter 3 Features and Functions
Chart Output (Analog Output)
98
Summary of Contents for 34420A
Page 1: ......
Page 2: ......
Page 13: ...1 Quick Start ...
Page 31: ...2 Front Panel Operation ...
Page 55: ...3 Features and Functions ...
Page 117: ...4 Remote Interface Reference ...
Page 200: ...5 Error Messages ...
Page 213: ...6 Application Programs ...
Page 245: ...6 Chapter 6 Application Programs Microsoft Excel Macro Example 245 ...
Page 246: ...Chapter 6 Application Programs Microsoft Excel Macro Example 246 ...
Page 247: ...7 Measurement Tutorial ...
Page 274: ...8 Specifications ...
Page 279: ...Dimensions Chapter 8 Specifications 34420A Nano Volt Micro Ohm Meter 280 ...
Page 294: ......