Setting the Function, Range, and Resolution
See also “General Measurement Configuration,” starting on page 55 in
chapter 3.
[SENSe:]FUNCtion "<
function>"
Select a measurement function. The function must be enclosed in quotes
in the command string (
for example, FUNC "VOLT:DC"
). Use one of
the following strings to set the function:
VOLTage:DC
RESistance (2-wire ohms)
VOLTage:DC:RATio
FRESistance (4-wire ohms)
VOLTage:DC:DIFFerence
TEMPerature
[SENSe:]FUNCtion?
This command queries the measurement function and return a quoted
string.
[{SENSe1: | SENSe2:}]VOLTage[:DC]:RANGe {<
range> | MIN | MAX}
[SENSe:]
FRESistance | RESistance
:RANGe {<
range> | MIN | MAX}
This command sets the range for the function specified. MIN selects the
lowest range.
MAX selects the highest range. [volatile memory]
[{SENSe1: | SENSe2:}]VOLTage[:DC]:RANGe? [MIN | MAX]
[SENSe:]
FRESistance | RESistance
:RANGe? {MIN | MAX}
This command queries the function range setting.
4
Chapter 4 Remote Interface Reference
Setting the Function, Range, and Resolution
137
Summary of Contents for 34420A
Page 1: ......
Page 2: ......
Page 13: ...1 Quick Start ...
Page 31: ...2 Front Panel Operation ...
Page 55: ...3 Features and Functions ...
Page 117: ...4 Remote Interface Reference ...
Page 200: ...5 Error Messages ...
Page 213: ...6 Application Programs ...
Page 245: ...6 Chapter 6 Application Programs Microsoft Excel Macro Example 245 ...
Page 246: ...Chapter 6 Application Programs Microsoft Excel Macro Example 246 ...
Page 247: ...7 Measurement Tutorial ...
Page 274: ...8 Specifications ...
Page 279: ...Dimensions Chapter 8 Specifications 34420A Nano Volt Micro Ohm Meter 280 ...
Page 294: ......