DC Voltage Noise Specifications *
Observation Period
Range
2-Minute
RMS
Noise
2-Minute
Peak-Peak
Noise
24-Hour
Peak-Peak
Noise
1 mV
1.3 nV
RMS
8 nVpp
12 nVpp
10 mV
1.5 nV
RMS
10 nVpp
14 nVpp
100 mV
10 nV
RMS
65 nVpp
80 nVpp
1 V
100 nV
RMS
650 nVpp
800 nVpp
10 V
450 nV
RMS
3
µ
Vpp
3.7
µ
Vpp
100 V
11
µ
V
RMS
75
µ
Vpp
90
µ
Vpp
* After a 2-hour warm-up,
±
1
°
C, 6.5 digits (10 PLC) with Analog
Filter Off and Digital Filter Medium (50 reading average). 2-Minute
RMS and 24-hour noise typical. For measurements using 0.02 or
0.2 NPLC, add 800 nV RMS noise.
DC Voltage Noise vs Source Resistance *
Source
Resistance
Noise
Analog Filter Digital Filter
0
1.3 nV
RMS
Off
Med
100
1.7 nV
RMS
Off
Med
1k
4 nV
RMS
Off
Med
10k
13 nV
RMS
Off
Med
100k
41 nV
RMS
On
Med
1M
90 nV
RMS
On
Slow
* Typical noise behavior for Ch 1 or Ch 2, after 2 hour warm-up,
6.5 digits (10 PLC), 2 minute observation period on 1 mV range.
For peak-to-peak noise, multiply RMS noise by 6.
MEASUREMENT CHARACTERISTICS
DC Voltage
Measurement Method
Continuously Integrating
Multi-slope III A-D Converter
A-D Linearity
0.00008% of r 0.00005% of range
Input Resistance
100V (Ch1 only)
1mV through 10V
10 M
Ω
±
1%
>10 G
Ω
, in parallel with <3.6 nF
Input Bias Current
<50 pA at 25
°
C
Injected Current*
<50 nA pp at 50 or 60 Hz
Input Protection
150 V peak any input terminal to
Channel 1 LO, continuous
Channel Isolation
Isolation between input channels
>10
10
Ω
Earth Isolation
350 V peak any input terminal to
earth. Impedence from any input
terminal to earth is
>10 G
Ω
and < 400 pF
Maximum Voltage
Channel 1 LO to Channel 2 LO,
150V peak
Resistance
Measurement Method
Selectable 4-wire or 2-wire ohms.
Current Source referenced to
Channel 1 LO input
Offset Compensation
Used on all ranges except
100 k
Ω
and 1 M
Ω
.
Can be turned off if desired
Protection
150V peak
Open Circuit Voltage
For Resistance and Low Power
Resistance <14 V. 20 mV, 100 mV,
500 mV selectable clamp
* Injected current is calibrated at the factory. Different line power
configurations will require recalibration.
8
Chapter 8 Specifications
34420A Nano Volt/Micro Ohm Meter
277
Summary of Contents for 34420A
Page 1: ......
Page 2: ......
Page 13: ...1 Quick Start ...
Page 31: ...2 Front Panel Operation ...
Page 55: ...3 Features and Functions ...
Page 117: ...4 Remote Interface Reference ...
Page 200: ...5 Error Messages ...
Page 213: ...6 Application Programs ...
Page 245: ...6 Chapter 6 Application Programs Microsoft Excel Macro Example 245 ...
Page 246: ...Chapter 6 Application Programs Microsoft Excel Macro Example 246 ...
Page 247: ...7 Measurement Tutorial ...
Page 274: ...8 Specifications ...
Page 279: ...Dimensions Chapter 8 Specifications 34420A Nano Volt Micro Ohm Meter 280 ...
Page 294: ......