To Make Null (Relative) Measurements
Each null measurement, also called
relative,
is the difference between a
stored null value and the input signal.
See page 77.
Reading
= measurement – null value
•
You can make null measurements on any function or channel. You can
set a null for each channel and function independently. The meter
remembers the null value when you switch between input channels or
functions.
•
To null the test lead resistance for more accurate two-wire ohms
measurements, short the ends of the test leads together and then
press
Null
.
•
The first reading taken after you press
Null
is stored as the null value
in the Null Register. Any previously stored value is replaced with the
new value.
•
After enabling null, you can edit the stored null value by
pressing
Shift >
(Menu Recall). This takes you to the “NULL
VALUE” command in the MATH MENU (
only if null is enabled
). Go
down to the “parameter” level, and then edit the displayed value.
Enables null operation;
Press again to disable
Null
annunciator is on when
null operation is enabled
2
Chapter 2 Front-Panel Operation
To Make Null (Relative) Measurements
45
Summary of Contents for 34420A
Page 1: ......
Page 2: ......
Page 13: ...1 Quick Start ...
Page 31: ...2 Front Panel Operation ...
Page 55: ...3 Features and Functions ...
Page 117: ...4 Remote Interface Reference ...
Page 200: ...5 Error Messages ...
Page 213: ...6 Application Programs ...
Page 245: ...6 Chapter 6 Application Programs Microsoft Excel Macro Example 245 ...
Page 246: ...Chapter 6 Application Programs Microsoft Excel Macro Example 246 ...
Page 247: ...7 Measurement Tutorial ...
Page 274: ...8 Specifications ...
Page 279: ...Dimensions Chapter 8 Specifications 34420A Nano Volt Micro Ohm Meter 280 ...
Page 294: ......