Agilent Technologies
3588A
Data Sheet
Product Specifications
Amplitude Specifications
All receiver specifications apply from 10 Hz to 150 MHz and include 30 minute
warmup from ambient conditions unless otherwise noted. Typical performance is
applicable over ±5°C from the most recent autocalibration and is not warranted.
Supplemental characteristics are non-warranted functional and feature
information.
Amplitude measurement range
Maximum Safe Input Level: 50/75 Ohm 1M Ohm
Average Continuous Power: 26 dBm - (10 Hz to 150 MHz)
Combined AC/DC ±4 Vpk ±25 Vpk
Maximum Measured Input 20 dBm
Input range settings (characteristic only)
50 Ohm input (in 10 dB steps) +20 dBm to -20 dBm
75 Ohm input (in 10 dB steps) +22 dBm to -18 dBm w/included adapter
1M Ohm input 0 dBm (for reference impedance of 50 Ohm)
A/D overload level:
2 dB (relative to selected range)
Dynamic range
Note: Dynamic range specifications apply with the source turned off.
Noise Level, 50 Ohm input (dBm/Hz using the marker noise function, -20 dBm
range)
The 50 ohm input noise level (dBm/Hz using the marker noise function,
-20 dBm range) will be typically -135dBc/Hz @ 1KHz and - 140 dBc/Hz