Scale fidelity

 (linearity)

Level Specified Typical

0 to -30 dB <0.05 dB 0.02 dB

-30 to -40 dB <0.1 dB 0.03 dB

-40 to -50 dB <0.3 dB 0.05 dB

-50 to -60 dB <0.5 dB 0.10 dB

-60 to -70 dB <0.7 dB 0.10 dB

-70 to -80 dB - 0.25 dB

-80 to -90 dB - 0.25 dB

-90 to -100dB - 0.40 dB

-100 to -110 dB - 0.70 dB

-110 to -120 dB - 4.0 dB

Level is relative to the input range
Specifications are valid for frequencies >100 kHz.

Example: To compute the typical accuracy for a signal of -45 dBm at
100 MHz with 50 Ohm full-scale range of -20 dBm and manual sweep,
sum the typical full-scale absolute accuracy and scale fidelity.

i.e. (0.2 dB + 0.02 dB) = 0.22 dB

Input Port Return Loss:

 >20 dB

Frequency Specifications

Frequency range

Specifications apply over the range of 10 Hz to 150 MHz
(10 Hz to 40 MHz for 1M Ohm input).

Frequency accuracy

Frequency accuracy is the sumofinitial accuracy, aging and
frequency counter resolution. Accuracy is measured with
frequency counter function.

Initial Accuracy Without opt. 001 With opt. 001

20 to 30°C ±0.5 ppm ±0.01 ppm

0 to 55°C ±3.0 ppm ±0.07 ppm

Summary of Contents for 3588A

Page 1: ...Ohm 1M Ohm Average Continuous Power 26 dBm 10 Hz to 150 MHz Combined AC DC 4 Vpk 25 Vpk Maximum Measured Input 20 dBm Input range settings characteristic only 50 Ohm input in 10 dB steps 20 dBm to 20...

Page 2: ...Spurious Unless specifically mentioned in other spurious specifications spurious responses are 70 dBc for signal levels range 80 dBc typical Harmonic distortion Low distortion mode 50 and 75 Ohm input...

Page 3: ...1 0 dB 0 5 dB 0 4 dB0 5 dB input 1 dB typ 0 5 dB typ 0 2 dB typ 0 2 dB typ 0 2 dB typ 75 Ohm 2 5 dB 1 0 dB 0 8 dB 0 8 dB 0 8 dB Input 1M Ohm 2 5 dB 1 0 dB 0 6 dB 0 6 dB Input The calibration procedur...

Page 4: ...curacy for a signal of 45 dBm at 100 MHz with 50 Ohm full scale range of 20 dBm and manual sweep sum the typical full scale absolute accuracy and scale fidelity i e 0 2 dB 0 02 dB 0 22 dB Input Port R...

Page 5: ...ncy span range characteristic only Swept spans Range 10 Hz to 150 MHz and zero span Resolution 0 1 Hz Accuracy Greater of 0 1 Hz or 125 of span Start stop freq 0 Hz to 150 MHz Narrowband zoom spans Ra...

Page 6: ...Hz s Narrowband zoom mode 7 measurements s for spans 10 kHz Narrowband zoom FFT 400 span Hz seconds time record length Note Traditional analog RBW filters are usually swept at RBW 2 Hz s or slower to...

Page 7: ...Non harmonic products 40 dB Noise 80dB Relative to the carrier in a 1 Hz bandwidth for offsets greater than 500 Hz from the carrier Output Port Return Loss 20 dB General Specifications Environmental...

Page 8: ...isk option Operating 4570m 15 000 ft Storage 4570m 15 000 ft Calibration Calibration Interval 1 year Warmup time 30 minutes Physical Power requirements 115 VAC operation 19 132 Vrms 47 440 Hz 230 VAC...