8-6
Agilent 4155C/4156C User’s Guide Vol.1, Edition 11
Specifications
Hardware
Table 8-2
Multiplier Factor of the Offset Current at 10 pA/100 pA Range
Table 8-3
HRSMU Voltage Range, Resolution, and Accuracy
Temperature
Humidity
5 to 60 % RH
60 to 80 % RH
5 to 18
°
C
×
2
×
2
18 to 28
°
C
×
1
×
2
28 to 40
°
C
×
2
×
5
Voltage
Range
Set
Reso.
Set Accuracy
a
a. Accuracy:
±
(% of reading or setting value + offset value).
Meas.
Reso.
Measurement
Accuracy
a
Max. Current
b
b. Vout is the output voltage in volts.
±
2 V
100
μ
V
±
(0.02 % + 400
μ
V)
2
μ
V
±
(0.01 % + 200
μ
V)
100 mA
±
20 V
1 mV
±
(0.02 % + 3 mV)
20
μ
V
±
(0.01 % + 1 mV)
100 mA
±
40 V
2 mV
±
(0.025 % + 6 mV)
40
μ
V
±
(0.015 % + 2 mV)
100 mA (Vout
≤
20 V)
50 mA
(20 V < Vout
≤
40 V)
±
100 V
5 mV
±
(0.03 % + 15 mV)
100
μ
V
±
(0.02 % + 5 mV)
100 mA (Vout
≤
20 V)
50 mA
(20 V < Vout
≤
40 V)
20 mA
(40 V < Vout
≤
100 V)
Summary of Contents for 41501A
Page 19: ...Contents Agilent 4155C 4156C User s Guide Vol 1 Edition 11 9 Accessories and Options...
Page 20: ...Agilent 4155C 4156C User s Guide Vol 1 Edition 11 Contents...
Page 21: ...1 Introducing the 4155C 4156C...
Page 41: ...2 Installation...
Page 84: ...2 44 Agilent 4155C 4156C User s Guide Vol 1 Edition 11 Installation Maintenance...
Page 85: ...3 Connecting to Network...
Page 129: ...4 File Operations...
Page 152: ...4 24 Agilent 4155C 4156C User s Guide Vol 1 Edition 11 File Operations Backing Up a Diskette...
Page 153: ...5 Print Plot Function...
Page 198: ...5 46 Agilent 4155C 4156C User s Guide Vol 1 Edition 11 Print Plot Function Output Examples...
Page 199: ...6 External Keyboard...
Page 203: ...7 Initial Settings...
Page 209: ...8 Specifications...
Page 243: ...9 Accessories and Options...