Step 1.
Make sure that the 42941A is correctly connected to the four-terminal pair port of the
Agilent 4294A.
Figure 4-8
Connecting the 42941A
NOTE
The 42941A must be directly connected to the four-terminal pair port of the Agilent
4294A. If another accessory such as a cable or adapter is inserted between the 42941A and
the four-terminal pair port of the Agilent 4294A, the correct adapter setup data cannot be
obtained and correct device measurement cannot be performed after the adapter setup.
Step 2.
More than 30 minutes warm-up time is required after turning on the Agilent 4294A.
Step 3.
Make the correct adapter selection for the 42941A in accordance with the “Adapter
Selection” on page 68.
Step 4.
After performing Step 4 of the “Adapter Selection” on page 68, press the
SETUP
key to
display the Adapter Setup Menu.
Step 5.
Leave the 3.5-mm port of the 42941A open (no connection). There is no OPEN standard
for the 42941A.
Step 6.
Press the
PHASE COMP [-]
key to start the phase compensation data measurement.
NOTE
The traces on the LCD display do not change during the phase compensation data
measurement.
When the phase compensation data measurement is completed, the softkey label changes
to
PHASE COMP [DONE]
.
Step 7.
With nothing connected to the 3.5-mm port, press the
OPEN [-]
key to start open data
measurement.
Summary of Contents for 4294A
Page 1: ......
Page 2: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 8: ......
Page 16: ...8 ...
Page 30: ...22 Chapter1 Installation Power Cable Figure 1 2 Alternative Power Cable Options ...
Page 70: ...62 Chapter3 Front Rear Panel and LCD Display Items Displayed on the LCD ...
Page 436: ...428 AppendixB Key Definitions Softkeys displayed by pressing the Recall key ...
Page 454: ...446 AppendixC Error messages WRONG I O PORT DIRECTION ...
Page 468: ......