Chapter 1 Performance Tests
Test 5: Power Meter Reference Level
Assembly Level Service Guide
1-37
1
19
Disconnect the Multimeter’s negative test lead from the 432A Power Meter’s
V
RF
connector, and connect it to the 432A Power Meter’s chassis ground.
20
Observe the reading on the Multimeter, and keep a record of this value
as V
COMP
.
21
Calculate the output level of the 53147A/148A/149A’s power reference
oscillator (Prf ) using the following formula:
Where:
Prf
=
2 V
comp
( V
1
– V
0
) + V
0
2
– V
1
2
4R (Calibration Factor)
Prf = Power reference oscillator output level
Vcomp = previously recorded value
V1 = previously recorded value
V0 = previously recorded value
R = previously recorded value
CalibrationFactor = value for thermistor mount at 50 MHz (traceable to NIST)