Chapter 2 Service
Calibration Procedures
2-12
Assembly Level Service Guide
2
Figure 2-3. Power Reference Oscillator Adjustment Locations
11
Set the frequency to 50.0 MHz ±0.05 MHz (49.95 MHz to 50.05 MHz).
Turn the adjustment tool counter-clockwise to increase the frequency or
clockwise to decrease the frequency.
R
98
(Power Adjustment)
L
9
(Frequency Adjustment)