Chapter 2 Operating Your Frequency Counter
Front Panel Display Messages
Operating Guide
2-63
2
Front Panel Display Messages
There are four types of displays:
•
Measurement Result Displays
•
Power-Up/Self Test Messages
•
Menu Messages
•
GPIB Messages
Measurement Result Displays
1
If this condition recurs, your Counter may need servicing.
Table 2-3. Measurement Result Displays
Display Message
Probable Cause
DOING STATS
You have chosen to see a statistics result and the Counter is calculating
statistics, but does not have valid statistics results yet.
1
INTERP FAIL
Interpolator calibration failed on the last measurement; therefore, no valid
measurement was taken.
1
MEAS FAIL
A measurement calibration failed on the last measurement.
NEW TIMEBASE
You have selected the auto reference mode (TIMEBAS: AUTO) and the Counter
detected that the external reference became invalid during the measurement.
Therefore, the current result is not valid, and the Counter switches to using the
internal reference.
NO TIMEBASE
You have selected external reference (TIMEBAS: EXT) and there is no external
reference applied to the rear-panel Ref In connector, or the external signal is not
an allowed frequency.
_ _ _ _ _ _ _ _ _ _ _ _
There is no valid measurement available to display. The Counter may be stopped
between measurements while in Single mode, and you need to press the
Stop/Single key; or trigger level or input conditioning may need adjustment.
Summary of Contents for 53181A
Page 2: ...NOTES...
Page 14: ...Contents x Operating Guide...
Page 24: ...Preface xx Operating Guide...
Page 33: ...1 Getting Started...
Page 68: ...Chapter 1 Getting Started Making Measurements 1 36 Operating Guide 1...
Page 69: ...2 Operating Your Frequency Counter Operator s Reference...
Page 143: ...3 Specifications...
Page 160: ...Chapter 3 Specifications Introduction 3 18 Operating Guide...