Chapter 2 Operating Your Frequency Counter
Front Panel Display Messages
2-64
Operating Guide
2
Power-Up/Self Test Messages
The following sequence occurs when you turn on the Counter:
1
All of the vacuum-fluorescent display segments and the indicators (or LEDs) are
illuminated for the power-up test.
2
If all tests pass:
SELFTST: PASS
will be displayed. (See NOTE below.)
3
HP-IB AT 3
is displayed, and then the Counter will start taking measurements.
NOTE
If there is a failure, each test that fails will display a failure message, and
SELFTST:
FAIL
will be displayed. You must press any key in order for the Counter to go on.
Possible individual failure messages are:
FAIL: DISP
display test
FAIL: CPU
processor test
FAIL: ROM
ROM test
FAIL: RAM
RAM test
FAIL: EEPROM
EEPROM test
FAIL: HP-IB
GPIB hardware test
FAIL: QSPI
serial hardware loops test
FAIL: FPGA
count circuit test
FAIL: FR END
front end test
FAIL: MEAS
measurement test
FAIL: INTERP
interpolator test
In addition, the following power-up message may be displayed:
UNCALIBRATED
This means that at least one type of calibration has NOT been performed.